Oxygen Deficiency Monitor

oxygen deficiency monitor

An oxygen deficiency monitor combines a room oxygen monitor and an oxygen depletion alarm. They are wall mounted indoors near stored gases like nitrogen, helium and argon. If one of the stored gases is released into the air, an oxygen depletion sensor inside will detect the lower oxygen level and activate an alarm to warn the occupants of potential oxygen deficiency in the room.

Why Are Oxygen Deficiency Monitors Important?

Oxygen deficiency monitors are important anywhere low oxygen levels in normal air can occur. 20.9% oxygen in the air is considered normal. When the oxygen level drops below 19.5% the air is considered oxygen-deficient. While it is still relatively safe for short periods of time, oxygen at levels below 16% is dangerous to human life.

The problem is that as the oxygen level drops, the volume of oxygen in the room is being replaced by gases like nitrogen, helium, argon or carbon dioxide. Because these gases are colorless and odorless, the oxygen deficiency may not be noticed immediately. However, if oxygen levels remain low, unconsciousness or death can occur.

Where are Oxygen Deficiency Monitors Used?

Oxygen deficiency monitors and alarms are necessary in any place where low oxygen levels can occur. The most common example is a room where pressurized cylinders or tanks of hydrogen, helium, argon, or nitrogen are utilized or stored. If one of the cylinders, tanks, hoses or fittings leak, the rapid depressurization of the gas storage system can lower the oxygen level in an enclosed area or room rapidly. Examples include:

  • Helium to cool MRI machines
  • Liquid nitrogen, helium or argon used for cryogenics
  • Liquid nitrogen to refrigerate package meats, fruits, or dairy
  • Industrial facilities that store argon and other shielding gases for welding

While carbon dioxide could be added to this list, common practice is to use a CO2 Safety Alarm to measure high CO2 levels instead of an oxygen deficiency alarm.

OSHA Requirements

OSHA's 29 CFR 1910.146: "Permit-Required Confined Spaces" details the requirements for employers to protect employees in areas that could potentially become oxygen deficient. Although the OSHA regulation primarily identifies areas like tanks, silos or pits that have limited access and are not designed for continuous employee occupancy, it is still good practice to protect employees who work around stored gases indoors where oxygen deficiency can occur.

The standard defines an oxygen-deficient atmosphere as any atmosphere containing less than 19.5% oxygen. Any area that contains less than 19.5% oxygen is hazardous and may not be entered by unprotected workers.

NIH Protocol for Oxygen Monitoring Devices

While not law, the National Institutes of Health Protocol for use and maintenance of Oxygen Monitoring Devices provides guidance on the installation, maintenance, and calibration of oxygen monitoring devices. It recommends:

“An oxygen monitoring device shall be installed in any indoor location where compressed gases and/or cryogenic liquids are stored and/or dispensed in manner that could create the potential for the displacement of oxygen. The NIH DRM notes that both “carbon dioxide manifold rooms… [and] nitrogen holding rooms shall include oxygen level monitoring alarms”. At a minimum, the following factors should be used in determining if a device should be installed: manufacturer (e.g., magnet) guidance, volume of gas used, work practices, location of gas, and ventilation estimates in the room/area. Additionally, compressed gases or cryogenic liquids shall not be located or dispensed in any indoor location that does not have adequate ventilation.”

Selecting an Oxygen Deficiency Monitor

When looking for an oxygen deficiency monitor, there are some features to consider:

  • Does it meet OSHA and NIH specifications? It should alarm at 19.5%, and 17.5%
  • Is it hard-wired? While wireless is easy to install, the sensor should be hard-wired to the remote display to guarantee that a WiFi or network failure does not disable the alarm.
  • Is it easy to install? Using an electrician to install the monitor will add significantly to the price.
  • Is it dependable? In most cases and oxygen deficiency monitor is an install once and forget it device. If the sensor fails it should inform you on the display.
  • Can it control the HVAC system? If you want the alarm to open a vent or turn on a fan to clear the air, does it have relays or industry standard outputs you can use?
  • Is it upgradeable? In noisy factory environments can strobes be added? Can you attach one or more remote displays to warn occupants of oxygen deficient air before they enter a room?

Where Should Oxygen Deficiency Alarms be Located?

All gases have molecular weight. Depending on the gas weight, as it displaces the oxygen it will tend to collect higher or lower in a room.

The molecular weight of air is 29. Therefore, gases that are much heavier should be mounted lower on the wall in a room while gases that are much lighter should be mounted higher. This insures the oxygen deficiency alarm’s sensor will sense the gas more quickly. Here’s a list of where to mount the oxygen sensor on a wall for the most common gases:

  1. Mount 18 inches from the floor: refrigerants, chlorine, butane, propane, CO2, nitrous oxide, argon.
  2. Mount 48 inches from the floor: ethylene, carbon monoxide, ammonia, nitrogen.
  3. Mount 60+ inches from the floor: hydrogen, helium.

Use this chart to find the molecular weight of the gas you are storing, then see if it is much lighter or heavier than air.

Oxygen Deficiency Monitors

oxygen deficiency monitor

Our RAD-0002 Oxygen Depletion Safety Alarm protects employees and customers near stored inert gases like nitrogen, argon, helium, nitrous oxide, welding gases and more. A gas leak can quickly displace oxygen in a room and make it unsafe. The RAD-0002 is used in hundreds of hospitals, MRI rooms, factories and cryogenic facilities nationwide.

This oxygen monitor combines a sensor unit and a wired remote display, both with audible and visual alarms. 3 built-in relays depending on the alarm level can be used to control an exhaust fan or send an alarm to the fire department or monitoring company. The alarm levels are user configurable to allow for specific applications.